Three Heating Zone Tube Furnace
|
-
Input power: 220VAC, 50/60Hz, single phase at Max. 5KW
-
Heating element: Fe-Cr-Al Alloy doped by Mo
-
1200oC Max. working temperature for < 60 minutes
-
1100oC Max. working temperature for continuous heating
-
Recommended heating rate: ≤ 10ºC/min
-
High purity quartz tube: φ100 x 1400mm
-
Total three heating zone length: 600mm (150mm + 300mm + 150mm)
-
Built-in PID Auto-Tune function with overheating & broken thermocouple protection
-
Temperature accuracy: ±1ºC
|
R2R Vacuum Chambers |
-
Two vacuum chambers are located in two ends of the furnace and connected with the quartz tube
-
The reeling rollers for a substrate in and out are located inside the vacuum chamber
-
Reeling speed for the substrate is adjustable from 1- 400mm/min
-
One roll of copper foil with 80 mm width and 0.025 mm thick (~2 kg) is included for customer practice
-
The max vacuum pressure for the chamber: 5 x 10^-2 torr
|
Three Channel Gas Delivery Station (Optional) |
-
Power: 220VAC, 50Hz, single-phase at Max. 23KW
-
Gas mixing tank: φ80 x 120mm
-
Working Temperature: 5~45ºC
-
Max. Pressure: 3 x 106 Pa
-
Accuracy: ±1.5% F.S.
-
Liner: ±(0.5-1.5)% F.S.
-
Duplicate accuracy: ±0.2%F.S.
-
Controlling Range:
-
Controller 1: 0~50 SCCM (Methane calibration)
-
Controller 2: 1~200 SCCM (H2 gas calibration)
-
Controller 3: 1~200 SCCM (Ar gas calibration)
-
Warning: H2 and Methane must be installed nearby the furnace to alarm gas leaking
|
Vacuum Pump and Valve
(Optional) |
-
Heavy-Duty Rotary Vane Vacuum Pump (7.8 CFM -240 L/m) with Two-Stage Exhaust System installed in the bottom case with max. vacuum pressure 10-2 Torr.
-
KFD25 adapter and stainless steel pipe are connected between the pump and tube flange with the ball valve
-
Digital vacuum pressure gauge and display are installed with the furnace
-
The oil-less pump and gas mixing & delivery system are available and PID pressure controlled vacuum system is available upon request at an extra cost
|
Anti-corrosive Pressure Gauge |
-
3.8 x 10-5 to 1125 Torr measurement range
-
Anti-corrosive, gas-type independent
-
High accuracy and reproducibility at atmosphere for reliable atmospheric pressure detection
-
Fast atmospheric detection eliminates waiting time and shortens process cycle
-
Easy to exchange plug & play sensor element
|
Temperature Controller |
-
Three temperature controllers with 30 programmable segments with +/- 1ºC temperature accuracy
-
Built-in PID Auto-Tune function with overheating & broken thermocouple protection
-
Over-temperature protection and alarm allow for operation without attendant(s)
-
RS485 Communications Port with PC operation software
-
You may upgrade the temperature controller to the Eurotherm Controller with 0.1ºC accuracy with software
|
Control Panel |
-
7" touch screen control panel which will control temperature, gas flow rate, plasma power
-
PC control software is available upon request for remote operation by Laptop computer (Must buy a laptop with pre-installed software)
|
Dimensions |
-
2400mm L x 600mm W x 1400mm H
|
Warranty |
One year limited warranty with lift time support (Consumable parts such as processing tubes, O-rings, and heating elements are not covered by the warranty, please order the replacement at related products page.) |